Abstract
This paper presents the application of the substitution method for the estimation of measurement uncertainties using calibrated workpieces in X-ray computed tomography (CT) metrology. We have shown that this, well accepted method for uncertainty estimation using tactile coordinate measuring machines, can be applied to dimensional CT measurements. The method is based on repeated measurements carried out on a calibrated master piece. The master piece is a component of a dose engine from an insulin pen. Measurement uncertainties estimated from the repeated measurements of the master piece were transferred on to additionally scanned uncalibrated workpieces which provided the necessary link for achieving traceable measurements.