Title of article :
Energy dispersive X-ray diffraction spectroscopy for rapid estimation of calcite in copper ores
Author/Authors :
Escلrate، نويسنده , , P. and Bailo، نويسنده , , D. and Guesalaga، نويسنده , , A. and Rossi Albertini، نويسنده , , V.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
6
From page :
566
To page :
571
Abstract :
This article presents a novel technique for measuring the concentration of calcite in copper ore samples using energy dispersive X-ray diffraction (ED-XRD). The results are compared with ultraviolet fluorescence spectroscopy (UVF) for the same group of samples. This approach substantially shortens the measuring time with respect to the current laboratory procedures from 300 min down to 15 min and 9 min for energy dispersive X-ray diffraction and ultraviolet fluorescence spectroscopy, respectively, while preserving a good accuracy level (RMSE < 0.577). dels used for data processing were calibrated and validated for both techniques via the correlation coefficients ( R c 2 and R v 2 ), the error of prediction (RMSEP) and the error of calibration (RMSEC). R v 2 values between 0.88 and 0.94 were obtained for UVF and ED-XRD data, respectively. The calibration and validation were performed for calcite concentration ranges of 3.4–10.7% in the case of UVF and 2.9–7.6% for ED-XRD. sults reported show the potential of this method for on-line estimation of calcite concentration.
Keywords :
Copper , diffraction , calcite , Spectroscopy , fluorescence
Journal title :
Minerals Engineering
Serial Year :
2009
Journal title :
Minerals Engineering
Record number :
2275527
Link To Document :
بازگشت