Title of article
MKIIIB-CTD: improving its system output
Author/Authors
Mِller، نويسنده , , Thomas J. and Holfort، نويسنده , , Jürgen and Delahoyde، نويسنده , , Frank and Williams، نويسنده , , Robert، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
14
From page
2113
To page
2126
Abstract
In the Neil Brown Instrumentsʹ MKIIIB-CTD (conductivity-temperature-depth profiler), the systemʹs digital outputs for the three basic measurements of temperature, conductivity and pressure typically show some small amplitude deviations from smooth calibrations which should be corrected for to achieve high accuracies, as required, e.g. within the Hydrographic Program (WHP) of the current World Ocean Circulation Experiment (WOCE). These deviations show up as (i) a strong nonlinearity or even discontinuity of several mK close to 0°C in temperature output leading to too high subzero temperatures; (ii) a jump of order 0.002 mS cm−1 in conductivity output when passing the half-range value 32.768 mS cm−1, which causes jumps in the relation of potential temperature and salinity; and (iii) errors in pressure measurements of up to 4 dbar due to mechanical hysteresis and both static and dynamic responses to temperature changes. The existence of these effects is demonstrated, and methods to reduce the associated errors are suggested.