• Title of article

    CHEMINI: A new in situ CHEmical MINIaturized analyzer

  • Author/Authors

    Vuillemin، نويسنده , , R. and Le-Roux، نويسنده , , D. and Dorval، نويسنده , , P. and Bucas، نويسنده , , K. and Sudreau، نويسنده , , J.P. and Hamon، نويسنده , , M. and Le Gall، نويسنده , , C. and Sarradin، نويسنده , , P.M.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    9
  • From page
    1391
  • To page
    1399
  • Abstract
    “CHEMINI” is a new instrument developed for the measurement of seawater chemical parameters. It is a mono-parameter in situ chemical analyzer based on flow injection analysis and colorimetric detection. The deep-sea version of CHEMINI combines two modules to perform the analysis of dissolved iron [Fe (II) or Fe (II+III)] and total sulphide (H2S+HS−+S2−) up to 6000 m depth. Detection limits are, respectively, 0.3 and 0.1 μM for iron and sulphide. The system proved highly reliable during the MoMARETO cruise on the Mid-Atlantic Ridge. The two CHEMINIs were used to describe the chemical environment in 12 mussel beds on the Tour Eiffel hydrothermal edifice.
  • Keywords
    Dissolved iron , hydrothermal vent , In situ analysis , Total sulphide
  • Journal title
    Deep Sea Research Part I: Oceanographic Research Papers
  • Serial Year
    2009
  • Journal title
    Deep Sea Research Part I: Oceanographic Research Papers
  • Record number

    2308834