Title of article :
Comparison between Amperometric and Chemiluminescence methods in detection and dosimetry of cobalt-60 gamma ray emission
Author/Authors :
Shourian، M. نويسنده Institute of Biochemistry and Biophysics, University of Tehran, P.O. Box 13145-1384, Tehran, Iran Shourian, M. , Ghourchian، H. نويسنده Laboratory of Microanalysis, Institute of Biochemistry & Biophysics, University of Tehran, Tehran, Iran. ,
Issue Information :
فصلنامه با شماره پیاپی 47 سال 2014
Abstract :
Background: Hydrogen peroxide (H2O2), produced by gamma ray irradiation to watery solution, was used as an analytical parameter for dosimetry of cobalt-60 (60Co) radiation. Materials and Methods: Detection of the produced H2O2 was carried out using two methods: an amperometric biosensor fabricated by immobilization of anthraquinone 2-carboxylic acid modified horseradish peroxidase on glassy carbon electrode, and chemiluminescence technique using luminol as a fluorophore and diperiodatocuprate as a catalyst. Results: In the first method, at the applied potential of -550 mV (vs. Ag/AgCl) the biosensor showed the sensitivity of 74.46 nA/Gray and detection limit of 0.061 Gray (R.S.D=3.67% for N=3) towards H2O2 produced by 60Co. In the second method, the sensitivity and detection limit for 60Co gamma ray was determined to be 2.34 (luminescence intensity/Gray) and 0.069 Gray (R.S.D=4.16% for N=3), respectively. Conclusion: The obtained results revealed that both proposed techniques, with leaner range from 0.25 to 5 Grays are applicable for detection and dosimetry of gamma ray from 60Co.
Journal title :
International Journal of Radiation Research
Journal title :
International Journal of Radiation Research