• Title of article

    Environmental stress cracking of polycarbonate catheter connectors

  • Author/Authors

    Lewis، نويسنده , , Peter R.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    9
  • From page
    1816
  • To page
    1824
  • Abstract
    A catheter connector allows medical staff to supply a range of fluids to ill patients using a single intravenous line. In the case of premature babies, such a line provides nutrition in the form of a synthetic milk (TPN) and antibiotics to prevent infection. The connectors are made from several alternative polymers, but polycarbonate is used widely. However, the material is susceptible to environmental stress cracking (ESC) from a variety of fluids. The case study shows how the first versions of one design were poorly moulded and cracked during service, allowing bacteria to enter the line and so infect the baby. The baby contracted meningitis, but lived although suffering brain damage. Evidence was produced which showed that it was a widespread problem in the mid-1990s in UK hospitals. The investigation suggested that poor design and manufacture were the root cause of the problem.
  • Keywords
    Polymer , Catheter , Connector , polycarbonate , fracture
  • Journal title
    Engineering Failure Analysis
  • Serial Year
    2009
  • Journal title
    Engineering Failure Analysis
  • Record number

    2338593