Title of article :
Neutronic calculations in support of the design of the ITER High Resolution Neutron Spectrometer
Author/Authors :
Moro، نويسنده , , F. and Esposito، نويسنده , , B. and Marocco، نويسنده , , D. and Villari، نويسنده , , R. and Petrizzi، نويسنده , , L. and Sunden، نويسنده , , E. Andersson and Conroy، نويسنده , , S. and Ericsson، نويسنده , , G. and Johnson، نويسنده , , M. Gatu and Dapena، نويسنده , , M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
5
From page :
1277
To page :
1281
Abstract :
This paper presents the results of neutronic calculations performed to address important issues related to the optimization of the ITER HRNS (High resolution Neutron Spectrometer) design, in particular concerning the definition of the collimator and the choice of the detector system. The calculations have been carried out using the MCNP5 Monte Carlo code in a full 3-D geometry. The HRNS collimation system has been included in the latest MCNP ITER 40° model (Alite-4). The ITER scenario 2 reference DT plasma fusion neutron source peaked at 14.1 MeV with Gaussian energy distribution has been used. n fluxes and energy spectra (>1 MeV) have been evaluated at different positions along the HRNS collimator and at the detector location. The noise-to-signal ratio (i.e. the ratio of collided to uncollided neutrons), the breakdown of the collided spectrum into its components, the dependency on the first wall aperture and the gamma-ray spectra at the detector position have also been analyzed. pact of the results on the design of the HRNS diagnostic system is discussed.
Keywords :
analysis , Spectrometer , ITER , Diagnostic , Neutron , Design
Journal title :
Fusion Engineering and Design
Serial Year :
2011
Journal title :
Fusion Engineering and Design
Record number :
2358140
Link To Document :
بازگشت