Title of article :
A study on the pulse height resolution of organic scintillator digitized pulses
Author/Authors :
Belli، نويسنده , , Francesco and Esposito، نويسنده , , Basilio and Marocco، نويسنده , , Daniele and Riva، نويسنده , , Marco، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Pages :
5
From page :
1271
To page :
1275
Abstract :
Organic scintillator detectors are widely used for neutron spectroscopy in fusion devices due to their good energy resolution and capability of neutron/gamma discrimination. Nowadays, scintillator pulses are commonly recorded by means of digital acquisition systems. These have several advantages, and in particular the possibility of off-line data reprocessing: however, the signal digitization can be a cause of degradation of the pulse height (PH) resolution (and therefore of the energy resolution obtained after pulse height spectra unfolding). In this work, the problem of how pulse digitization may influence the pulse height resolution is investigated. First, through downsampling of digitized scintillator pulses, we determine the minimum sampling rate required to avoid any degradation of the pulse height resolution. Secondly, we find that the application of interpolation methods to the digitized pulses does not affect the pulse height resolution, whereas it increases the figure of merit for neutron/gamma discrimination. These results are relevant to define the specifications for the digital acquisition systems of neutron detectors in present and future fusion devices such as JET and ITER.
Keywords :
Neutron/gamma discrimination , Digital acquisition , Pulse height resolution , Interpolation , Organic scintillator
Journal title :
Fusion Engineering and Design
Serial Year :
2013
Journal title :
Fusion Engineering and Design
Record number :
2361262
Link To Document :
بازگشت