Title of article :
Implementation of a quality management system at the PHOENIX facility (CryoMaK)
Author/Authors :
Urbach، نويسنده , , Elisabeth and Bagrets، نويسنده , , Nadezda and Weiss، نويسنده , , Klaus-Peter، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Abstract :
Within a variety of mechanical tests in the Cryogenic Material Test Facility Karlsruhe (CryoMaK) at Karlsruhe Institute of Technology (KIT) the PHOENIX facility was prepared for multiple standard tensile tests in liquid helium, liquid nitrogen and at room temperature. With the multiple specimens holder 10 specimens can be tested within one cool down one after another. A quality management system is needed for ensuring reproducible preconditions. For the guarantee of the competence of the laboratory and the measurement equipment, a quality management system was implemented and prepared for accreditation according to DIN EN ISO/IEC 17025 (ISO 17025). The implementation of a quality management system allows high precision test results included the estimation of measurement uncertainty. This paper gives an overview of the management and technical requirements for the accreditation of the PHOENIX testing facility.
Keywords :
CryoMaK , accreditation , measurement uncertainty , tensile test , ISO 17025