• Title of article

    Evaluation of thermal load during laser corneal refractive surgery using infrared thermography

  • Author/Authors

    Brunsmann، نويسنده , , U. and Sauer، نويسنده , , U. and Arba-Mosquera، نويسنده , , S. and Magnago، نويسنده , , T. and Triefenbach، نويسنده , , N.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    6
  • From page
    342
  • To page
    347
  • Abstract
    Infrared thermography is used for evaluation of the mean temperature as a measure of thermal load during corneal refractive surgery. An experimental method to determine emissivity and to calibrate the thermografic system is presented. In a case study on the porcine eye two dimensional temperature distributions with lateral resolution of 170 μm and line scans with temporal resolution of 13 μs are discussed with respect to the meaning of mean temperature. Using the newest generation of surgery equipment it is shown, that the mean temperature rise can be kept below 5 °C during myopic laser in situ keratomileusis (LASIK) treatments corresponding to an aberration-free correction of −2.75 diopter.
  • Keywords
    thermography , Laser corneal refractive surgery , thermal load , emissivity
  • Journal title
    Infrared Physics & Technology
  • Serial Year
    2010
  • Journal title
    Infrared Physics & Technology
  • Record number