Title of article :
Design and thermal analysis of electrically calibrated pyroelectric detector
Author/Authors :
Shao، نويسنده , , Xiumei and Ding، نويسنده , , Jieying and Ma، نويسنده , , Xueliang and Yu، نويسنده , , Yuehua and Fang، نويسنده , , Jiaxiong، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
4
From page :
45
To page :
48
Abstract :
A finite element model was set up by ANSYS to optimize the structure of an electrically calibrated detector and simulate the thermal diffusivity. High-performance relaxor-based ferroelectric crystals were selected as the detector materials. Gold-black was selected as the black coating on the surface of the calibrated pyroelectric detector. The finite element model was set up by ANSYS and the geometry model was meshed appropriately. The optimized structure stood out from the several feasible structures. The simulation results showed that the differences of ΔT on the top surface of PMN-PT crystal between radiant and electrical heating decrease with the increase of sensitive area. The sensitive area of the detector is 10 mm × 10 mm and the entrance aperture is ∅ 8 mm. In this condition, the differences of ΔT are 1.20%. An electrically calibrated pyroelectric detector was fabricated according to the simulation results.
Keywords :
Finite element model , thermal analysis , Electrically calibrated pyroelectric detector
Journal title :
Infrared Physics & Technology
Serial Year :
2012
Journal title :
Infrared Physics & Technology
Record number :
2376025