Title of article :
Determining the Effective Source-Surface Distance for Therapeutic Electron Beams
Author/Authors :
Tahmasebi Birgani، Mohammad Javad نويسنده Department of Medical Physics, School of Medicine, Ahvaz
Jundishapur University of Medical Sciences, Ahvaz, IR
Iran , , Zabihzadeh، Mansour نويسنده Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, IR Iran , , Arvandi، Sholeh نويسنده Department of Clinical Oncology, Golestan Hospital, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, IR Iran , , Gharibreza، Elahe نويسنده Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, IR Iran ,
Issue Information :
فصلنامه با شماره پیاپی 29 سال 2016
Abstract :
Background: The source point of the irradiated electron beam must be considered to estimate the output factor and dose distribution
during electron therapy.
Objectives: The aim of this study was to determine the effective source-surface distance (SSDeff) of an electron linear accelerator
(Linac), and its dependence on energy and depth.
Materials and Methods: A Varian Linac 2100CD with electron energies of 4, 6, 9, 12, and 15 MeV, electron applicator size of 2020
cm2, nominal SSDs of 97 to 113 cm, and air gaps of 2 to 18 cm were studied. Using a Farmer (0.13 cc) ionizing chamber, the percentage
depth doses were measured in the water phantom (50 cm3) and then the SSDeff was calculated by applying the inverse square law.
Results: For the 100% PDD, the SSDeff values were calculated as 79, 91, 92, 93, and92cmfor 4, 6, 9, 12, and15MeV, respectively. At a depth
with a certain PDD, increasing energy also increases SSDeff, and a similar increase is observed at a distinctive energy by increasing
the PDD.
Conclusions: Using the maximum dose depth from PDD curves and the inverse square law, the required SSDeff to calculate the dose
distribution of the electron beam can be calculated.