Title of article :
Descemetʹs Stripping Automated Endothelial Keratoplasty Tissue Insertion Devices
Author/Authors :
Nasir Khan، Salman نويسنده , , Shiakolas، Panos S نويسنده , , Vinod Mootha، Venkateswara نويسنده ,
Issue Information :
فصلنامه با شماره پیاپی سال 2015
Abstract :
This review study provides information regarding the construction, design, and use of six commercially
available endothelial allograft insertion devices applied for Descemet’s stripping automated endothelial
keratoplasty (DSAEK). We also highlight issues being faced in DSAEK and discuss the methods through
which medical devices such as corneal inserters may alleviate these issues. Inserter selection is of high
importance in the DSAEK procedure since overcoming the learning curve associated with the use of an
insertion device is a time and energy consuming process. In the present review, allograft insertion devices
were compared in terms of design, construction material, insertion technique, dimensions, incision
requirements and endothelial cell loss to show their relative merits and capabilities based on available data
in the literature. Moreover, the advantages/disadvantages of various insertion devices used for allograft
insertion in DSAEK are reviewed and compared. The information presented in this review can be utilized
for better selection of an insertion device for DSAEK.
Keywords :
Allograft Insertion Device , Allograft Inserter , Descemets Stripping Endothelial Keratoplasty , Corneal Injectors , Descemets Stripping Automated Endothelial Keratoplasty