Title of article :
Partial Coherence Laser Interferometry in Highly Myopic versus Emmetropic Eyes
Author/Authors :
Roessler, Gernot F Department of Ophthalmology - RWTH Technical University of Aachen, Aachen, Germany , Djalali Talab, Yassin Department of Ophthalmology - RWTH Technical University of Aachen, Aachen, Germany , Dietlein, Thomas S Department of Ophthalmology - University of Cologne, Cologne, Germany , Dinslage, Sven Department of Ophthalmology - University of Cologne, Cologne, Germany , Plange, Niklas Department of Ophthalmology - RWTH Technical University of Aachen, Aachen, Germany , Walter, Peter Department of Ophthalmology - RWTH Technical University of Aachen, Aachen, Germany , Mazinani, Babac AE Department of Ophthalmology - RWTH Technical University of Aachen, Aachen, Germany
Pages :
5
From page :
169
To page :
173
Abstract :
Purpose: To investigate the reliability of partial coherence laser interferometry for optical biometry in highly myopic eyes. Methods: Axial length measurements by the IOLMaster (Carl Zeiss Meditec, Germany) with signal-to-noise ratio (SNR) ≥2 were performed in 52 consecutive myopic subjects with axial length ≥26.5 mm and 45 emmetropic patients before cataract surgery. Axial length measurements and SNR were analyzed and compared among the two study groups. Results: Axial length measurements were feasible in 46 of 52 (88.5%) highly myopic eyes and in 41 of 45 (91.1%) eyes with normal axial length. To achieve two reliable axial length values with SNR ≥2, a mean number of 2.06±0.25 measurements was necessary in myopic eyes and 2.10±0.37 in emmetropic counterparts. Mean SNR after two measurements was 4.98±2.44 in myopic eyes versus 5.56±2.32 in control eyes. Even though successful measurement was independent of preoperative visual acuity, patients with visual acuity better than 20/63 showed significantly higher SNR values. Conclusions: Partial coherence laser interferometry shows satisfying feasibility and good signal quality for axial length determination in highly myopic eyes with stable retinal condition and clear media.
Keywords :
Axial Length , Partial Coherence Interferometry , High Myopia , Biometry
Journal title :
Astroparticle Physics
Serial Year :
2014
Record number :
2422440
Link To Document :
بازگشت