Author/Authors :
Menezes, Sílvio Emanuel Acioly Conrado de Dental School, University of Pernambuco FOP-UPE - Tabatinga - Camaragibe-PE, Brazil , Batista, Shirley Machado Dental School, University of Pernambuco FOP-UPE - Tabatinga - Camaragibe-PE, Brazil , Lira, Juliana Ouro Preto Dental School, University of Pernambuco FOP-UPE - Tabatinga - Camaragibe-PE, Brazil , Monteiro, Gabriela Queiroz de Melo Dental School, University of Pernambuco FOP-UPE - Tabatinga - Camaragibe-PE, Brazil
Abstract :
Introduction: Endodontic instruments are developed to provide a better cleaning of the root
canal system and reduce its risk of fracture. The aim of this study was to evaluate the
instrumentation time and cyclic fatigue resistance of WaveOne Gold, ProDesign R and
ProDesign Logic files. Methods and Materials: Thirty Nickel-titanium (NiTi) rotary
instruments were divided into 3 groups (n=10). ProDesign Logic file 25/0.06 was used in
continuous rotation after glide path preparation. WaveOne Gold 25/0.07 and ProDesign R
25/0.06 files were used in reciprocating motion. Every file instrumented 3 standardized
artificial canals. The average time, the number of cycles (NCI) and cyclic fatigue resistance of
each file were determined through the number of cycles to failure (NCF) in a stainless-steel
device. The total amount of cycles to fracture was also calculated (NCI+NCF). Data was
analyzed using the Kruskal-Wallis and Mann-Whitney tests. Results: The instrumentation
time of the ProDesign Logic file was significantly lower when compared to the other files
(P=0.019). The longest times to failure were presented by ProDesign Logic (182.07 sec) and
ProDesign R (152.38 sec) files. The same differences were observed for the NCF (910.37 and
761.93). The WaveOne Gold group presented a lower NCF as well as a smaller sum of
NCI+NCF (748.33) that was statistically significant when compared to the other groups
(P<0.05) respectively. Conclusion: The use of continuous rotational motion in canals with a
glide path in the ProDesign Logic group led to shorter instrumentation time. The cyclic
fatigue resistance of ProDesign R and Logic instruments was superior to WaveOne Gold. The
thermal treatment of the instrument’s alloy, its cross section and the glide path seems to
influence the cyclic fatigue resistance.
Keywords :
Cyclic Fatigue , ProDesign Logic , ProDesign R , WaveOne Gold