Author/Authors :
Tahmasebi Birgani, Mj Radiation Therapy Department - Golestan Hospital - Joundi Shapoor University ofMedical Sciences , Karbalaee, SM Radiation Therapy Department, Ahwaz
Abstract :
Background: In external radiation therapy, the percentage depth dose (PDD) is an important factor for estimating
patient dose and dose distribution in target volume; therefore, its exact measurement or calculation is important.
The aim of this study was to evaluate analytically the dose received by different points in water phantom
and to compare it with dosimetry measurement data.
Methods: To find the dose distribution throughout the tumor volume, first, the mathematical approach was performed
for derivation of percentage depth dose photon beams of 6MV and 18MV Varian accelerator. Second, by
dosimetry for different fields in different depths of water phantom, one can parameterize the obtained formula for
percentage depth dose.
Results: By comparing the mathematical and dosimetry results, the parameters of PDD-expression were computed
in terms of the dimension of equivalent square field in different depths. From this formula, one can find the
PDD for any fields in different depths, surface skin dose, and depth of build-up region of dose distribution, which
are in agreement with empirical results with R2 >0.995, showing a good agreement with the experimental data.
Conclusion: So one can measure the surface skin-dose, the depth of build-up region, and their variations in
terms of square field size exactly; the measurement of these quantities have some technical problems in radiation
dosimetry.
Keywords :
Dosimetry , Surface dose , Build-up region , Radiation therapy