Author/Authors :
Borhani, Setareh Department of Chemical Engineering - School of Chemical and Petroleum Engineering - Shiraz University , Hassanajili, Shadi Department of Nanochemical Engineering - School of New Science and Technology - Shiraz University , Ahmadi Tafti, Seyed Hossein Research Center for Advanced Technologies in Cardiovascular Medicine - Tehran Heart Center - Tehran University of Medical Sciences , Rabbani, Shahram Research Center for Advanced Technologies in Cardiovascular Medicine - Tehran Heart Center - Tehran University of Medical Sciences
Abstract :
Compared to bare-metal stents (BMSs), drug-eluting stents (DESs) have been regarded as a revolutionary change in coronary
artery diseases (CADs). Releasing pharmaceutical agents from the stent surface was a promising progress in the realm of
cardiovascular stents. Despite supreme advantages over BMSs, in-stent restenosis (ISR) and long-term safety of DESs are
still deemed ongoing concerns over clinically application of DESs. The failure of DESs for long-term clinical use is associated
with following factors including permanent polymeric coating materials, metallic stent platforms, non-optimal drug
releasing condition, and factors that have recently been supposed as contributory factors such as degradation products of
polymers, metal ions due to erosion and degradation of metals and their alloys utilizing in some stents as metal frameworks.
Discovering the direct relation between stent materials and associating adverse effects is a complicated process, and yet it
has not been resolved. For clinical success it is of significant importance to optimize DES design and explore novel strategies
to overcome all problems including inflammatory response, delay endothelialization, and sub-acute stent thrombosis (ST)
simultaneously. In this work, scientific reports are reviewed particularly focusing on recent advancements in DES design
which covers both potential improvements of existing and recently novel prototype stent fabrications. Covering a wide
range of information from the BMSs to recent advancement, this study mostly sheds light on DES’s concepts, namely stent
composition, drug release mechanism, and coating techniques. This review further reports different forms of DES including
fully biodegradable DESs, shape-memory ones, and polymer-free DESs.
Keywords :
Coronary artery diseases , Cardiovascular stents , In-stent restenosis , Stent thrombosis