• Title of article

    A New Method for Measurement of Radio Frequency Inhomogeneity in MRI

  • Author/Authors

    Oghabian, M. A Tehran University of Medical Sciences - Medical Physics department, Tehran , Riahi Alam, N Tehran University of Medical Sciences - Medical Physics department, Tehran , Mehdipour, S Imam Hospital, Tehran

  • Pages
    7
  • From page
    113
  • To page
    119
  • Abstract
    Weintroducea newmethodofmeasuringtransmitandreceiveRFinhomogeneity indifferentRFcoils ofMRIsystems. In thismethoda single sliceofa uniformphantom is imagedfromdifferentflipangles,usinga standardspinechoprotocol.Thesignalintensityin these images is then fitted to a mathematicalmodelwhich describes the relationshipbetween thesignalintensityandflipangleofthespinechoimages.Theresultsofthiscurve fittingprocess aretwoparameters,T(r)and R(r),whosevariationwith thespatialposition showsRFtransmit,andreceivenon-uniformity,respectively. Inthisapproacha linearprofileofB1fielddistributionandreceivesensitivityofRFcoilsare achievedwhichisalsoapplicableinvivo. Themethodcanbeusedtoassessanycommercial MR scannerand is highly recommended for thequality control (QC) of thoseMRscanners whicharedevoidofcomplicatedprotocolssuchasSE(-2).Suchsystemsarestillrunning indifferentclinicsespeciallyinthedevelopingcountrieswherethelatesthighperformance MRscannersarenotavailableandmanyscannerslackstandardmaintenanceservices.
  • Keywords
    Magnetic Resonance Imaging , RF inhomogeneity , Image nonuniformity , spin echo
  • Journal title
    Astroparticle Physics
  • Serial Year
    2003
  • Record number

    2434871