    Zamany, J Cardiovascular Research Center - Faghihi Hospital - Shiraz University of Medical Sciences, Shiraz, Iran , Mahmoody, Y Cardiovascular Research Center - Faghihi Hospital - Shiraz University of Medical Sciences, Shiraz, Iran , Ostavan, MA Cardiovascular Research Center - Faghihi Hospital - Shiraz University of Medical Sciences, Shiraz, Iran

  • Pages
    3
  • From page
    139
  • To page
    141
  • Abstract
    A 63-year-old woman with typical chest pain referred to our center. Her risk factors were hypertension (HTN), hyperlipidemia (HLP), diabetic mellitus (DM), and smoking. Coronary angiography (CAG) revealed a tubular lesion (95%) in proximal part, and a diffuse lesion (95%) in the mid part of left anterior descending (LAD) with a good distal flow. Stenting of proximal and mid part lesion was done for her with 2 drug-eluting stent (paclitaxel- eluting coronary stent system, EURoCOR) with success. After 6 days, due to a typical chest pain without response to medications, the coronary angiography was performed showing a definite fracture of stent which was implanted in the mid part of left anterior descending artery (LAD) 6 days before admission.
  • Keywords
    Drug-eluting stent , Fracture , Implantation