Title of article :
Contralateral assessment of sub-Bowman keratomileusis (SBK) microkeratome suction duration on laser-assisted in-situ keratomileusis (LASIK) flap characteristics
Author/Authors :
Yousif, Mohamed Omar Ophthalmology Department - Faculty of Medicine, Ain Shams University, Egypt , Abdelfattah, Nizar Saleh Doheny Eye Institute - University of California Los Angeles, USA , Zayed, Amira A. Department of Surgery - Mayo Clinic, Rochester, USA , Soliman, Ashraf H. Ophthalmology Department - Faculty of Medicine, Ain Shams University, Egypt
Pages :
4
From page :
65
To page :
68
Abstract :
Purpose: To describe the effect of prolonging the standard suction duration during laser-assisted in-situ keratomileusis (LASIK) and its effect on flap thickness and hinge length using sub-Bowman keratomileusis (SBK) microkeratome. Methods: Fifty-six eyes (28 patients) were included and divided into 2 groups; Group-A: eyes with flatter corneas (36 eyes, 18 patients) and mean keratometric readings ranging from 40.13 to 43.71 diopters (D). Group-B: eyes with steeper corneas (20 eyes, 10 patients) with mean keratometric readings ranging from 43.85 to 46.72 D. One-Use-Plus SBK microkeratome was used for flap creation. For right eyes, flap was created immediately once suction was built up. In left eyes, the surgeon waited for 10 s after suction was built up before flap creation. Flap hinge length and flap thickness were measured using surgical caliper and ultrasonic pachymetry, respectively. Results: Statistically significant differences were observed in corneal flap hinge size between right eyes versus left eyes, with a mean of 3.98 ± 0.48 vs. 3.78 ± 0.55 mm ( p < 0.001). Mean flap thickness in both eyes did not prove to be statistically significantly different with either surgical technique (90.2 ± 1.68 vs. 90.07 ± 1.44 mm, p ¼ 0.8). Sub-group analysis of Group-A vs. Group-B revealed hinge sizes that were significantly larger in steeper corneas (p < 0.01 and p < 0.05, respectively). However, flap thickness in both groups was unaffected by surgical procedure (p ¼ 0.5). Conclusions: Increasing suction duration increases flap hinge length and stabilizes the flap, especially in steeper corneas.
Keywords :
Laser-assisted in-situ keratomileusis , LASIK; Suction , Corneal flap
Journal title :
Journal of Current Ophthalmology
DOI :
Serial Year :
2016
Journal title :
Journal of Current Ophthalmology
Record number :
2444644