• Title of article

    Implantation of multiple suprachoroidal electrode arrays in rabbits

  • Author/Authors

    Abdallah, Walid Department of Ophthalmology - USC Roski Eye Institute - Keck School of Medicine of the University of Southern California, USA , Li, Wen Department of Electrical and Computer Engineering - Michigan State University, East Lansing, MI, USA , Weiland, James Department of Ophthalmology - USC Roski Eye Institute - Keck School of Medicine of the University of Southern California, USA , Humayun, Mark Department of Ophthalmology - USC Roski Eye Institute - Keck School of Medicine of the University of Southern California, USA , Ameri, Hossein Department of Ophthalmology - USC Roski Eye Institute - Keck School of Medicine of the University of Southern California, USA

  • Pages
    6
  • From page
    68
  • To page
    73
  • Abstract
    Purpose: Epiretinal and subretinal prosthesis have been shown to be a valid way to provide some vision to patients with advanced outer retinal degeneration and profound vision loss. However, the field of vision for these patients is markedly limited by the area occupied by the electrode array. In this study, we aimed to evaluate the feasibility of implantation of multiple suprachoroidal electrode arrays in a single eye in order to increase the field of vision in patients implanted with retinal prosthesis. Methods: The right eye of seventeen Dutch rabbits (age range, 5e6 months) was used for the study. Multiple inactive custom-made electrode arrays were inserted into the suprachoroidal space (SCS) and animals were followed up for up to 6 months using fundus photography, optical coherence tomography (OCT), and fluorescein angiography (FA). Results: It was possible to surgically implant up to 8 electrode arrays in a single eye. None of the rabbits showed any major complications. The electrodes were well tolerated and remained in position in all rabbits. There was no evidence of retinal damage on follow-up exams and FA throughout the study. Conclusion: Multiple suprachoroidal electrode array implantation is feasible and may provide a novel approach to increase the field of vision in subjects implanted with retinal prosthesis.
  • Keywords
    Suprachoroidal , Electrodes , Array , Retinal prosthesis , Retinal implant
  • Journal title
    Journal of Current Ophthalmology
  • Serial Year
    2018
  • Journal title
    Journal of Current Ophthalmology
  • Record number

    2447741