Author/Authors :
Malmir, Maryam Faculty of Chemistry - Bu-Ali Sina University, Hamedan, Iran , Afkhami, Abbas Faculty of Chemistry - Bu-Ali Sina University, Hamedan, Iran , Madrakian, Tayyebeh Faculty of Chemistry - Bu-Ali Sina University, Hamedan, Iran , Soltani-Felehgari, Farzaneh Faculty of Chemistry - Bu-Ali Sina University, Hamedan, Iran , Madrakian, Elaheh Faculty of Chemistry - Bu-Ali Sina University, Hamedan, Iran
Abstract :
An anodic stripping differential pulse voltammetric method was proposed for rapid and sensitive determination of hypochlorite (OCl-)
ion. A modified electrode was prepared by modification of carbon paste using alumina nanopowder. This rapid, simple and accurate
method is based on the electrooxidation of OCl- ion accumulated at the electrode surface. Nanoalumina modification increased the
oxidation peak current for OCl- ion. The effects of different parameters on the electrode response were studied and the optimum condition
was established. The response of the sensor was linear in the range 0.1-800 μg ml-1 of OCl- ion, with a correlation coefficient of 0.9992 at
the optimum condition. The limit of detection was obtained as 0.025 μg ml-1. The effects of some cations, anions and organic species,
which may coexist with OCl- ion in real samples on the current response of OCl-, were investigated. The investigated chemical species did
not interfere with the OCl- electrochemical signal. OCl- ion in water and dairy product samples was successfully determined using the
proposed electrode.
Keywords :