Title of article :
X-ray spectra calculation for different target-filter of mammograms using MCNP Code
Author/Authors :
Mowlavi, A.A Tarbiat Moallem University of Sabzevar
Pages :
5
From page :
129
To page :
133
Abstract :
Background: An electron beam generated X-ray spectrum consists of characteristic X-ray and continuous bermsstrahlung. The aim of this research is calculating and comparing X-ray spectra for different target filter of mammograms. Materials and Methods: Monte Carlo is a very powerful tool to simulate a series of different target-filter assembly in order to calculate the X-ray spectra. MCNP version 4C has been used for simulation set up to calculate X-ray spectra for different target-filter combination of mammograms. Results: The spectra of different and the most commonly used target– filter in mammography equipments (Mo-Mo, Mo-Rh, Rh-Rh, W-Rh, Mo-Al and Rh-Al) have been calculated. Conclusion: The computational results can be used to select a suitable X-ray spectrum for mammography as well as for computing the absorbed dose. Also, the Monte Carlo result of Mo-Mo is in good agreement with the published results.
Keywords :
Monte Carlo simulation , mammography , target- filter combination , X-ray spectra , MCNP code
Journal title :
Astroparticle Physics
Serial Year :
2005
Record number :
2473261
Link To Document :
بازگشت