• Title of article

    Detection of lead ions using an electrochemical aptasensor

  • Author/Authors

    Adabi ، Mohsen - Islamic Azad University, Roudehen Branch

  • Pages
    6
  • From page
    247
  • To page
    252
  • Abstract
    The aim of this research was to develop a simple, selective and sensitive aptasensor for detection of Pb2+ based on hairpin structure of complementary strand (CS) of aptamer (Apt). Screen printed carbon electrode (SPCE) was electrodeposited by gold nanoparticles to be used as the substrate for the immobilization of AptCS. Moreover, gold nanoparticles (AuNPs) and thionine were added to increase the sensitivity of the aptasensor. The morphology of SPCE and AuNPs modified SPCE was investigated using scanning electron microscopy (SEM). The SEM results showed that the AuNPs were homogeneously distributed on the surface of SPCE. The electrochemical performance of aptasensor was evaluated by means of cyclic voltammetry (CV) and differential pulse voltammetry (DPV) measurements. The electrochemical results indicated that peak current due to the hairpin structure formation of CS decreased in the presence of Pb2+. However, in the absence of Pb2+, peak current was enhanced because of the combination of thionine/AuNPsCS. The designed aptasensor showed a high selectivity toward Pb2+, wide linear range (140 nM), low detection limit (374 pM), acceptable reproducibility and good longterm stability.
  • Keywords
    Aptasensor , Cyclic voltammetry , Differential pulse voltammetry , Lead ions , Gold nanoparticles
  • Journal title
    Nanomedicine Research Journal
  • Serial Year
    2019
  • Journal title
    Nanomedicine Research Journal
  • Record number

    2478743