• Title of article

    Intrapelvic Protrusion of a Broken Guide Wire Fragment during Fixation of a Femoral Neck Fracture

  • Author/Authors

    Saini, Uttam Chand Department of Orthopedics - Postgraduate Institute of Medical Education and Research, Chandigarh, India , Gopinathan, Nirmal Raj Department of Orthopedics - Postgraduate Institute of Medical Education and Research, Chandigarh, India , Aggarwal, Sameer Department of Orthopedics - Postgraduate Institute of Medical Education and Research, Chandigarh, India , Sen, Ramesh Department of Orthopedics - Postgraduate Institute of Medical Education and Research, Chandigarh, India

  • Pages
    2
  • From page
    155
  • To page
    156
  • Abstract
    Breakage of DHS guide wire during surgery and its migration into the pelvis through the hip joint is a rare complication and its removal can be very challenging for the surgeon. We share our experience of a similar case wherein we used an ‘iliofemoral’ approach to successfully remove the broken transfixing guidewire from the hip joint. Although iliofemoral approach is similar to the lateral window of conventional ilioinguinal approach, yet it is less invasive, has lesser complications, requires less expertise and is easily reproducible by an average orthopaedic trauma surgeon. We recommend that surgical approaches for removal of these broken or migrated wires should be individualized depending upon the exact location of the wire tip in the hip joint or pelvis and need for exposure.
  • Keywords
    Guide wire , Dynamic hip screw , Proximal hip fracture , Iliofemoral approach , complication
  • Journal title
    Astroparticle Physics
  • Serial Year
    2018
  • Record number