Title of article
Analysis of Hydroxyl Methyl Furfuran by Graphene Oxide-reinforced Hollow Fiber Electromembrane Extraction in Food Samples
Author/Authors
Rezaee ، Maryam Department of Chemistry - Islamic Azad University, Mashhad Branch , Shoeibi ، Shahram Food and Drug Laboratory Research Center, Iran Food and Drug Administration (IFDA) - Ministry of Health and Medical Education (MOH) , Ebrahimi ، Mahmoud Department of Chemistry - Islamic Azad University, Mashhad Branch
Pages
9
From page
155
To page
163
Abstract
Cooking, sterilization and roasting are used in the food industry for different purposes, such as taste improvement, flavor addition, and shelf-life extension. The roasting stage plays an important role in the improvement of color and sensory properties of coffee; although as a result of this process the level of Hydroxymethylfurfural (HMF) is increased. The purpose of the this investigation was to analysis the level of HMF in instant coffee and infant formula samples through Electromembrane extraction (EME) techniques coupled with HPLC. To achieve the highest extraction efficiency, some critical extraction parameters were optimized, including the concentration of geraphen oxide (GO), the pH of the sample solution, the solution existing inside the lumen of hollow fiber, the extraction technique, and desorption time. A full validation was conducted for the method in terms of linearity, precision, and trueness. The method showed good linear ranges (0.1^-100μg kg^-1) with correlation coefficients greater than 0.999- 0.998 μg kg^-1 for instant coffee and infant formula. The limit of quantification (LOD) value of the present method for instant coffee and infant formula was 0.1μg kg^-1. The method provided high recoveries ranging from 85.0% to 108.0% for instant coffee and infant formula. The results indicated that the method was rapid, efficient and environmentfriendly in the determination of HMF in instant coffee and infant formula samples.
Keywords
Graphene oxide , Hollow fiber , Hydroxymethylfurfural , High , performance liquid chromatography , Electromembrane techniques
Journal title
Journal of Chemical Health Risks
Record number
2494503
Link To Document