Title of article :
Diagnostic reference levels at intraoral and dental panoramic examinations
Author/Authors :
Hodolli, G Institute of Occupational Medicine - Radiation Protection Service, 15000 Kastriot, Kosovo , Kadiri, S Institute of Occupational Medicine - Radiation Protection Service, 15000 Kastriot, Kosovo , Nafezi, G Institute of Occupational Medicine - Radiation Protection Service, 15000 Kastriot, Kosovo , Bahtijari, M University of Prishtina “HasanPrishtina” - Department of Physics, 10000 Pristine, Kosovo , Syla, N University of Prishtina “HasanPrishtina” - Department of Physics, 10000 Pristine, Kosovo
Abstract :
Background: The aim of this study was to measure and determine diagnostic
reference levels (DRLs) of programs used at intraoral and dent panoramic
examinations in Kosovo. Materials and Methods: To determine DRL, were
performed measurements for dose area product (DAP) at intraoral and dental
panoramic radiology systems. This study has covered over 47% of X-ray units, which
are in use for intraoral and panoramic imaging examinations on the country.
Results: From results, we conclude that, the mean of DAP (mGy cm2) for
single intraoral examination is 26.8, 29.6 and 39.7 for incisor, premolar and
molar, respectively. Therefore the mean of DAP (mGy cm2) value for
panoramic dental radiology is 62.7, 74.1 and 90.3 for child, adult and large
adult, respectively. For each modality mention above are proposed specific
DRLs. Conclusion: Recommended DRLs value for intraoral radiography for
incisor, premolar and molar programs are: 28.5, 36.5 and 50.3 mGy cm2,
respectively. Also, DRLs for dental panoramic radiography for child, adult and
large adult programs are: 73.0, 81.0 and 93.0 mGy cm2, respectively.
A better quality assurance (QA) and quality control (QC) should be
enforcement. For strengthening of QA and QC engagement of radiographers
and Medical Physicist is mandatory.
Keywords :
dent , panoramic , intraoral , dose area product , Reference levels