Title of article :
Dosimetric Accuracy Comparison between ACUROSE XB, AAA and PBC Dose Calculation Algorithms in EclipseTM TPS Using a Heterogeneous Phantom
Author/Authors :
Rooshenass, Raheleh Department of Medical Radiation Engineering - Central Tehran Branch - Islamic Azad University , Salehi Barough, Mehdi Department of Medical Radiation Engineering - Central Tehran Branch - Islamic Azad University , Gholami, Somayeh Department of Radiotherapy - Cancer Institute - School of Medicine - Tehran University of Medical Sciences , Mohammadi, Ehsan Department of Radiotherapy - Cancer Institute - School of Medicine - Tehran University of Medical Sciences
Pages :
6
From page :
168
To page :
173
Abstract :
Purpose: The aim of this study is to compare the accuracy of different algorithms in EclipseTM Treatment Planning System (TPS) using a heterogeneous phantom. Materials and Methods: The method is based on the International Atomic Energy Agency's TEC-DOC 1583 report. The chest phantom of CIRS, PTW30010 ionization chamber and an electrometer (PTW, Freiburg) were used for the measurements. Three ACUROSE XB (AXB), Analytical Anisotropic Algorithm (AAA) and Pencil Beam Convolution (PBC) dose calculation algorithms available in Eclipse TM TPS were considered in this study. Results: Based on the measurements, the maximum differences between calculated dose by TPS and measured dose in TEC-DOC 1583 tests were 2.5%, 8.6% and 16.1% for the AXB, AAA and PBC algorithms in heterogeneous media, respectively. Conclusion: The Acuros XB algorithm has superior accuracy to predict the dose distribution in the heterogeneous tissues such as lung compared to AAA and PBC algorithms.
Keywords :
Heterogeneous Tissues , Treatment Planning System , Dose Calculations
Journal title :
Frontiers in Biomedical Technologies
Serial Year :
2019
Record number :
2500504