Title of article :
Determination of Uterus Absorbed Dose by Patients following Myocardial Perfusion Scan using TLD and Conjugate View Methods
Author/Authors :
Shanei, A Department of Medical Physics - School of Medicine - Isfahan University of Medical Sciences, Isfahan, Iran , Heydari, F Department of Medical Physics - School of Medicine - Isfahan University of Medical Sciences, Isfahan, Iran , Moslehi, M Department of Medical Physics - School of Medicine - Isfahan University of Medical Sciences, Isfahan, Iran
Abstract :
Introduction: The determination of patient’s absorbed dose is the first step
of radiation protection which depends on the quantification of organ activity in
nuclear medicine. The aim of the present study was to determine the absorbed dose
by patient’s uterus following myocardial perfusion scan with 99mTC-sestamibi using
Theroluminescnce dosimetry (TLD) and conjugate-view methods.
Material and Method: In this study, each patient was injected 15 to 20 mCi
(based on their weight) of 99mTC-sestamibi. Myocardial perfusion scan from twenty
two patients (females) were acquired by gamma camera at 15, 60 and 90 minutes
after 99mTC-sestamibi injection. To determine the amount of activity in uterus, conjugate
view method was applied on images. Then, MIRD equation was used to estimate
absorbed dose in uterus of patients. Moreover, uterus absorbed dose was determined
using TLD method. At the end, absorbed dose values obtained in conjugate view
method were compared with the data obtained from TLD method.
Results: The average amount of uptake for 99mTC-sestamibi by heart was calculated
3.077±0.067 percent of injected dose. The uterus activity at the intervals of 15,
60 and 90 minutes after injection of 99mTC-sestamibi was 0.044±0.015, 0.031±0.014
and 0.026±0.013 mCi, repectively. The uterus absorbed dose per unit of injected
activity (mGy/MBq × 10−4) obtained 5.258±0.500 using TLD method.
Conclusion: The results of this study were in good agreement with similar studies.
Dosimetry using TLD, in comparison with the conjugate view method, demonstrates
more accurate results.
Keywords :
Uterus , Thermoluminescent Dosimetry , Myocardial Perfusion Imaging , Radionuclide Imaging
Journal title :
Journal of Biomedical Physics and Engineering