Title of article
Measurement of dielectric loss tangent at cryogenic temperature using superconducting film resonator
Author/Authors
Zhang, Yufang College of Physics and Electronic Engineering - Northwest Normal University - Lanzhou 730070 - Gansu, China , Wang, Zhenqing College of Physics and Electronic Engineering - Northwest Normal University - Lanzhou 730070 - Gansu, China
Pages
6
From page
27
To page
32
Abstract
We demonstrate that the superconducting film resonator can be used to accurately and quantitatively measure the microwave dielectric loss tangent of a variety of materials. Compared to traditional dielectric resonator
loaded metal cavity method, it has advantage of small
sample size (*2–3 orders of magnitude smaller than the
old method), and much higher sensitivity to measure small
loss tangent values as small as 10-5 at around 7 GHz band
at cryogenic temperatures. This method can be utilized
widely in study of mechanism of microwave loss at cryogenic
temperature range, which is extremely important in
superconducting microwave application areas, such as
novel super quantum computers.
Keywords
Dielectric loss tangent , Superconducting plate resonator , Cryogenic temperature
Journal title
Journal of Theoretical and Applied Physics
Serial Year
2016