Xu, Manman School of Mechanical and Automotive Engineering - Anhui Polytechnic University - Wuhu 241000, China - Institute of Plasma Physics - Chinese Academy of Sciences - Hefei 230026, China , Song, Yuntao Institute of Plasma Physics - Chinese Academy of Sciences - Hefei 230026, China , Chen, Gen Institute of Plasma Physics - Chinese Academy of Sciences - Hefei 230026, China , Li, Hengbo Hefei CAS Ion Medical and Technical Devices Co - Ltd - Hefei 230088, China , Wang, Jun Hefei CAS Ion Medical and Technical Devices Co - Ltd - Hefei 230088, China
Abstract :
The superconducting cyclotron for proton therapy that is under design by ASIPP (Hefei, China) and JINR (Dubna, Russia)
will be able to accelerate protons to the energy 200 MeV with the maximum beam current of 1 lA. The performance of
superconducting magnets is the key to the superconducting cyclotron, and it will directly affect the quality of the proton
beam and the therapeutic effect. To measure the magnetic induction intensity of the superconducting coil with the
cryogenic Hall probes on the testing platform. As the current has been increased from 0 to 160 A, the measurement error of
probes decreases obviously whose tendencies are consistent. By comparing the measuring results with the theoretical
results, we can confirm the quality of superconducting coil and the installation location. Meanwhile, the basic operation of
cryogenic Hall probes and data acquisition system is grasped, and the experimental experience for the magnetic intensity
test of superconducting cyclotron has been accumulated.