Title of article :
Comparison the Accuracy of Fetal Brain Extraction from T2- Half-Fourier Acquisition Single-Shot Turbo Spin-Echo (HASTE) MR Image with T2-True Fast Imaging with Steady State Free Precession (TRUFI) MR Image by Level Set Algorithm
Author/Authors :
Zabihzadeh, Mansour Department of Medical Physics and Department of Clinical Oncology - Faculty of Medicine - Golestan Hospital - Ahvaz Jundishapur University of Medical Sciences, Iran , Pishghadam, Morteza Department of Medical Physics - Faculty of Medicine - Ahvaz Jundishapur University of Medical Sciences, Iran , Kazemi, Kamran Department of Electrical and Electronics Engineering - Shiraz University of Technology, Iran , Nekooei, Sirous Department of Radiology - Faculty of Medicine - Mashhad University of Medical Sciences, Iran , Tahmasebi Birgani, Mohamad Javad Department of Medical Physics and Department of Clinical Oncology - Faculty of Medicine - Golestan Hospital - Ahvaz Jundishapur University of Medical Sciences, Iran , Seilanian-Toosi, Farrokh Department of Radiology - Faculty of Medicine - Mashhad University of Medical Sciences, Iran
Abstract :
Background
Access to appropriate images of fetal brain can greatly assist to diagnose of probable abnormalities.
The aim of this study was to compare the suitability of T2-True Fast Imaging with Steady State Free
Precession (T2-TRUFI), and T2-Half-Fourier Acquisition Single-Shot Turbo Spin-Echo (T2-
HASTE( magnetic resonance imaging (MRI) to extract the fetal brain using the level set algorithm.
Materials and Method: T2-TRUFI and T2-HASTE MRI of the uterus were performed. The fetal
brain was cropped from the image manually, with an adequate margin of maternal tissues; and then
the fetal brain was extracted using level set. The outcome was statistically analyzed to compare its
success, error (Sensitivity and Specificity), and similarities (Dice and Jaccard), with those of images
obtained by radiologist.
Results: The mean values of statistical tests to evaluate the similarity (Dice and Jaccard) and the
success and error (Sensitivity and specificity) between two T2-TRUFI and T2-HASTE were
calculated as 97.35%, 94.98%, 95.88%, 95.88%, 99.45%, 91.10%,83.82%, 86.44% and 99.11%,
respectively. However, the results from two images showed high scores to extract the fetal brain, but
images from the T2-HASTE technique resulted in better visually output.
Conclusion
Based on our results, the T2-HASTE MR images are preferred to fetal brain extraction by level set
algorithm compared to the T2-TRUFI MR images.
Keywords :
Level set algorithm , Fetal brain extraction , MR images , Sensitivity , Specificity
Journal title :
International Journal of Pediatrics