Title of article :
Portable system of programmable syringe pump with potentiometer for determination of promethazine in pharmaceutical applications
Author/Authors :
Saleh, Tawfik A. King Fahd University of Petroleum Minerals - Chemistry Department,, Saudi Arabia , Abulkibash, A.M. King Fahd University of Petroleum Minerals - Chemistry Department, Saudi Arabia , Ibrahim, Atta E. University of Khartoum - Faculty of Pharmacy - Department of Pharmaceutical Chemistry, Sudan
From page :
155
To page :
160
Abstract :
A simple and fast-automated method was developed and validated for the assay of promethazine hydrochloride in pharmaceutical formulations, based on the oxidation of promethazine by cerium in an acidic medium. A portable system, consisting of a programmable syringe pump connected to a potentiometer, was constructed. The developed change in potential during promethazine oxidation was monitored. The related optimum working conditions, such as supporting electrolyte concentration, cerium(IV) concentration and flow rate were optimized. The proposed method was successfully applied to pharmaceutical samples as well as synthetic ones. The obtained results were realized by the official British pharmacopoeia (BP) method and comparable results were obtained. The obtained t-value indicates no significant differences between the results of the proposed and BP methods, with the advantages of the proposed method being simple, sensitive and cost effective
Keywords :
Portable syringe pump , Potentiometry , Promethazine , Pharmaceutical applications
Journal title :
Saudi Pharmaceutical Journal(SPJ)
Journal title :
Saudi Pharmaceutical Journal(SPJ)
Record number :
2552881
Link To Document :
بازگشت