• Title of article

    Determination of Fano Factor and Pre-amplifier Noise from the Measurement of Energy Resolution of a HPGe Detector

  • Author/Authors

    SAMAT, S. B. Universiti Kebangsaan Malaysia - School of Applied Physics, Faculty of Science and Technology, Malaysia , PRIHARTI, W. Universiti Kebangsaan Malaysia - School of Applied Physics, Faculty of Science and Technology, Malaysia

  • From page
    761
  • To page
    764
  • Abstract
    A HPGe detector and γ-ray multinuclide standard solution were used to get several data pairs of photon energy E and detector’s energy resolution FWHM. A series of graph of FWHM2 versus E was plotted to determine the detector’s Fano factor and pre-amplifier noise, which yielded the mean values of 0.123±0.003 and (7.448±0.011) × 10^−1 keV, respectively. The Fano factor value agrees within 95% confident interval with the mostly quoted value of 0.13 for semiconductor detectors. The obtained preamplifier noise is in agreement with typical manufacturer’s specification.
  • Keywords
    Fano factor , FWHM , gamma , ray spectrometry , pre , amplifier noise
  • Record number