• Title of article

    Pacemaker Twiddler s Syndrome

  • Author/Authors

    ZAMIRIAN, M. shiraz university of medical sciences - Division of Cardiology - Department of Internal Medicine,, شيراز, ايران , SHARIF KAZEMI, M. B. shiraz university of medical sciences - Division of Cardiology - Department of Internal Medicine,, شيراز, ايران , MOARREF, A. R. shiraz university of medical sciences - Division of Cardiology - Department of Internal Medicine,, شيراز, ايران

  • From page
    59
  • To page
    60
  • Abstract
    A 76-yr-old man with unipolar ventricular pacemaker, previously implanted in the right ventricular apex, presented with syncopal attack and twitching of the right pectoralis muscle. He finally was diagnosed as having Twiddler s syndrome. The patient had a pulse rate of 36/min and cannon waves in the neck veins. Chest radiography showed a dislodged pacing lead completely wrapped around the pulse generator. After uncoiling the lead it was positioned in the right ventricular apex and the pulse generator implanted and sewed tightly in a smaller pocket. After six months follow-up there was no further twiddling or pacemaker dysfunction
  • Keywords
    Twiddler s syndrome • pacemaker • syncopal attacK
  • Journal title
    Iranian Journal of Medical Sciences (IJMS)
  • Journal title
    Iranian Journal of Medical Sciences (IJMS)
  • Record number

    2571064