Title of article
Computer Vision Based Automatic Extraction and Thickness Measurement of Deep Cervical Flexor from Ultrasonic Images
Author/Authors
Kim, Kwang Baek Department of Computer Engineering - Silla University - Busan, Republic of Korea , Song, Doo Heon Department of Computer Games - Yong-in Songdam College - Yongin, Republic of Korea , Park, Hyun Jun Department of Computer Engineering - Pusan National University - Busan, Republic of Korea
Pages
11
From page
1
To page
11
Abstract
Deep Cervical Flexor (DCF) muscles are important in monitoring and controlling neck pain. While ultrasonographic analysis
is useful in this area, it has intrinsic subjectivity problem. In this paper, we propose automatic DCF extractor/analyzer software
based on computer vision. One of the major difficulties in developing such an automatic analyzer is to detect important organs
and their boundaries under very low brightness contrast environment. Our fuzzy sigma binarization process is one of the answers
for that problem. Another difficulty is to compensate information loss that happened during such image processing procedures.
Many morphologically motivated image processing algorithms are applied for that purpose. The proposed method is verified as
successful in extracting DCFs and measuring thicknesses in experiment using two hundred 800 × 600 DICOM ultrasonography
images with 98.5% extraction rate. Also, the thickness of DCFs automatically measured by this software has small difference (less
than 0.3 cm) for 89.8% of extracted DCFs.
Keywords
Cervical , Flexor , Measurement , Ultrasonic , DCF
Journal title
Computational and Mathematical Methods in Medicine
Serial Year
2016
Full Text URL
Record number
2607412
Link To Document