Title of article
Efficacy of Amflow®, a Real-Time-Portable Feedback Device for Delivering Appropriate Ventilation in Critically Ill Patients: A Randomised, Controlled, Cross-Over Simulation Study
Author/Authors
Kim, Jong Won Department of Emergency Medicine - Konkuk University Medical Center - Konkuk University School of Medicine - Seoul - Republic of Korea , Park, Sang O Department of Emergency Medicine - Konkuk University Medical Center - Konkuk University School of Medicine - Seoul - Republic of Korea , Lee, Kyeong Ryong Department of Emergency Medicine - Konkuk University Medical Center - Konkuk University School of Medicine - Seoul - Republic of Korea , Hong, Dae Young Department of Emergency Medicine - Konkuk University Medical Center - Konkuk University School of Medicine - Seoul - Republic of Korea , Baek, Kwang Je Department of Emergency Medicine - Konkuk University Medical Center - Konkuk University School of Medicine - Seoul - Republic of Korea
Pages
7
From page
1
To page
7
Abstract
Objective. ,e aim of this study was to test whether Amflow® (a newly designed portable ventilation feedback device) can assist rescuers in delivering target tidal volume (VT) and respiration rate (RR) during self-inflating bag (SB) ventilations in various clinical scenarios. Method. ,is was a simulation study with a prospective cross-over design. A total of 40 trained participants who underwent
training for SB ventilation were recruited. Using a SB with or without Amflow® alternately, participants delivered ventilations to test
lungs connected to a gas flow analyser in each of three different scenarios: acute respiratory distress syndrome (ARDS; 315–385 ml
ranges for 350 ml target VT, with 20 breaths/min); cardiopulmonary resuscitation (CPR; 450–550 ml ranges for 500 ml target VT with
10 breaths/min); and adult head trauma (630–770 ml ranges for 700 ml target VT with 15 breaths/min). Results. ,e feedback group
(SB with Amflow®) demonstrated a significantly higher percentage of delivering the appropriate VT ranges than the no-feedback
group for both ARDS (58.6% versus 23.5%, respectively) and CPR (85.4% versus 41.0%, respectively) (all p < 0.05). However, there was
no significant difference between the two groups in the percentage of delivering the appropriate VT ranges in head trauma patients
(65.9% versus 68.3%, respectively; p � 0.092). In all three scenarios, a higher percentage of target RR delivered was achieved in the
feedback group (88.3%, 99.2%, and 96.3%, respectively) compared with the no-feedback group (5.8%, 12.5%, and 10.0%, respectively)
(all p < 0.05). Conclusion. ,e Amflow® device could be useful for rescuers in delivering SB ventilation with appropriate VT and RR simultaneously in various critical situations, except for clinical cases that demand greater delivered VT.
Keywords
test whether Amflow® , CPR , Critically Ill Patients , Randomised , Controlled
Journal title
Emergency Medicine International
Serial Year
2020
Record number
2607457