Title of article
Novel Use of an Orbital Atherectomy Device for In-Stent Restenosis: Lessons Learned
Author/Authors
Shaikh, K. USD Department of Medicine - Division of Cardiology, Sioux Falls, USA , Kelly, S. USD Department of Medicine - Division of Cardiology, Sioux Falls, USA , Gedela, M. USD Department of Medicine - Division of Cardiology, Sioux Falls, USA , Kumar, V. USD Department of Medicine - Division of Cardiology, Sioux Falls, USA , Stys, A. USD Department of Medicine - Division of Cardiology, Sioux Falls, USA , Stys, T. USD Department of Medicine - Division of Cardiology, Sioux Falls, USA
Pages
5
From page
1
To page
5
Abstract
We present a case of a 67-year-old man with stage III chronic kidney disease, uncontrolled diabetes mellitus, coronary artery disease,
and high surgical risk who presented with two episodes of acute coronary syndrome attributed to in-stent restenosis (ISR) associated
with heavily calcified lesions. In this case, we were able to improve luminal patency with orbital atherectomy system (OAS); however,
withdrawal of the device resulted in a device/stent interaction, causing failure of the device. Given limitations in current evidence
and therapies, managing ISR can be a technical and cognitive challenge. Balloon expansion of the affected region often provides
unsatisfactory results, possibly related to significant calcium burden. OAS could be an efficacious way of reestablishing luminal
patency in ISR lesions, as these lesions are often heavily calcified.
Keywords
Orbital Atherectomy Device , In-Stent Restenosis
Journal title
Case Reports in Cardiology
Serial Year
2016
Full Text URL
Record number
2607879
Link To Document