Title of article
A Rare Case of Complete Fragmentation of Pacemaker Lead after a High-Velocity Theme Park Ride
Author/Authors
Khattak, Furqan East Tennessee State University, Johnson City, TN, USA , Khalid, Muhammad East Tennessee State University, Johnson City, TN, USA , Gaddam, Sathvika East Tennessee State University, Johnson City, TN, USA , Ramu, Vijay East Tennessee State University, Johnson City, TN, USA , Brahmbhatt, Vipul East Tennessee State University, Johnson City, TN, USA
Pages
5
From page
1
To page
5
Abstract
Pacemaker lead fracture is one of the most common causes of pacemaker malfunction and is most frequently associated with weight
lifting or chest trauma. These patients usually present with symptoms of dizziness, syncope, chest discomfort, and palpitations or
less commonly with extracardiac symptoms. Diagnosis is made by ECG and careful review of chest imaging such as chest X-ray or
fluoroscopy. Treatment involves placement of a new lead with or without extraction of the fractured lead. We present an interesting
case of complete severance of the tip of a dual-chamber pacemaker’s atrial lead after a high-velocity theme park ride. In our case, the
fracture occurred during amusement park rides and went undiagnosed until the patient presented for routine pacemaker
evaluation. This case indicates that extreme physical forces in the absence of direct trauma, such as during amusement park
rides, may result in lead fractures and patients with pacemakers should be cautioned regarding such activities.
Keywords
Complete Fragmentation , Pacemaker Lead , High-Velocity Theme Park Ride
Journal title
Case Reports in Cardiology
Serial Year
2018
Full Text URL
Record number
2611098
Link To Document