• Title of article

    Transcutaneous Puncture of an Undeflatable Coronary Angioplasty Balloon Catheter

  • Author/Authors

    Leibundgut, Gregor Department of Cardiology - Medizinische Universitätsklinik, Rheinstrasse, Switzerland , Degen, Christian Departement of Internal Medicine - Medizinische Universitatsklinik, Liestal, Switzerland , Riede, Florian Department of Cardiology - Medizinische Universitätsklinik, Rheinstrasse, Switzerland

  • Pages
    6
  • From page
    1
  • To page
    6
  • Abstract
    This case report describes a quick and safe method to successfully retrieve an undeflatable angioplasty balloon via the transradial access site. The article also presents potential mechanisms of contrast media entrapment and bench tests of guidewire-assisted balloon puncture. After successful stent implantation in the proximal right coronary artery in a 74-year-old female patient referred for acute coronary syndrome, the balloon catheter became undeflatable for an unknown reason. Several attempts to burst the balloon in the guide catheter and the aorta failed. After a pullback into the radial artery, the still inflated balloon became stuck again and was unable to be retrieved through the sheath. Finally, transcutaneous puncture and aspiration of contrast media with a 23 gauge syringe needle through the skin in the right forearm allowed deflation and successful removal the balloon.
  • Keywords
    Undeflatable Coronary , Angioplasty Balloon Catheter
  • Journal title
    Case Reports in Cardiology
  • Serial Year
    2018
  • Record number

    2611129