Title of article :
Fluid Structure Interaction on Paravalvular Leakage of Transcatheter Aortic Valve Implantation Related to Aortic Stenosis: A Patient-Specific Case
Author/Authors :
Basri, Adi A Department of Aerospace Engineering - Universiti Putra Malaysia - Serdang, Malaysia , Zuber, Mohammad Department of Aeronautical and Automobile Engineering - Manipal Institute of Technology - Manipal Academy of Higher Education - Manipal, India , Basri, Ernnie I Department of Aerospace Engineering - Universiti Putra Malaysia - Serdang, Malaysia , Zakaria, Muhammad S Fakulti Kejuruteraan Mekanikal - Universiti Teknikal Malaysia Melaka - Hang Tuah Jaya - Durian Tunggal - Melaka, Malaysia , Aziz, Ahmad F. A Department of Medicine - Universiti Putra Malaysia - Serdang, Malaysia , Tamagawa, Masaaki Department of Biological Function and Engineering - Kyushu Institute of Technology - Kitakyushu, Japan , Ahmad, Kamarul A Department of Aerospace Engineering - Universiti Putra Malaysia - Serdang, Malaysia
Abstract :
*is study investigated the impact of paravalvular leakage (PVL) in relation to the different valve openings of the transcatheter
aortic valve implantation (TAVI) valve using the fluid structure interaction (FSI) approach. Limited studies were found on the
subject of FSI with regards to TAVI-PVL condition, which involves both fluid and structural responses in coupling interaction.
Hence, further FSI simulation with the two-way coupling method is implemented to investigate the effects of hemodynamics
blood flow along the patient-specific aorta model subjected to the interrelationship between PVL and the different valve openings
using the established FSI software ANSYS 16.1. A 3D patient-specific aorta model is constructed using MIMICS software. *e
TAVI valve identical to Edward SAPIEN XT 26 (Edwards Lifesciences, Irvine, California), at different Geometrical Orifice Areas
(GOAs), is implanted into the patient’s aortic annulus. *e leaflet opening of the TAVI valve is drawn according to severity of
GOA opening represented in terms of 100%, 80%, 60%, and 40% opening, respectively. *e result proved that the smallest
percentage of GOA opening produced the highest possibility of PVL, increased the recirculatory flow proximally to the inner wall
of the ascending aorta, and produced lower backflow velocity streamlines through the side area of PVL region. Overall, 40% GOA
produced 89.17% increment of maximum velocity magnitude, 19.97% of pressure drop, 65.70% of maximum WSS magnitude, and
a decrement of 33.62% total displacement magnitude with respect to the 100% GOA.
Keywords :
Patient-Specific , PVL , TAVI , GOA , Paravalvular