Author/Authors :
Ruiz, Oliver Grimaldo Department of Structural - Geotechnical and Building Engineering (DISEG) - Politecnico di Torino - Turin, Italy , Dhaher, Yasin Physical Medicine and Rehabilitation - Orthopaedic Surgery - UT Southwestern Medical Center - Dallas - TX , USA
Abstract :
This study reports on a new method for the development of multi-color and multi-material realistic
Knee Joint anatomical models with unique features. In particular, the design of a fibers matrix structure that mimics
the soft tissue anatomy.
Methods: Various Computer-Aided Design (CAD) systems and the PolyJet 3D printing were used in the fabrication
of three anatomical models wherein fibers matrix structure is mimicked: (i) Anterior cruciate ligament reconstruction
(ACL-R) model used in the previous study. (ii) ACL-R model, incorporating orientations, directions, locations, and
dimensions of the tunnels, as well as a custom-made surgical guide (SG) for avoiding graft tunnel length mismatch.
(iii) Total knee arthroplasty (TKA) model, including custom-made implants. Before models 3D printing, uni-axial
tensile tests were conducted to obtain the mechanical behaviors for individual No. 1 (A60-A50), No. 2 (A50-A50),
No. 3 (A50-A40), and No. 4 (A70-A60) soft tissue-mimicking polymers. Each material combination represents
different shore-hardness values between fiber and matrix respectively.
Results: We correlated the pattern of stress-strain curves in the elastic region, stiffness, and elastic modulus of
proposed combinations with published literature. Accordingly, material combinations No. 1 and No. 4 with elastic
modules of 0.76-1.82 MPa were chosen for the soft tissues 3D printing. Finally, 3D printing Knee Joint models were
tested manually simulating 50 flexo-extension cycles without presenting ruptures.
Conclusion: The proposed anatomical models offer a diverse range of applications. These may be considered as an
alternative to replacing cadaver specimens for medical training, pre-operative planning, research and education
purposes, and predictive models validation. The soft tissue anatomy-mimicking materials are strong enough to
withstand the stretching during the flexo-extension. The methodology reported for the design of the fiber-matrix
structure might be considered as a start to develop new patterns and typologies that may mimic soft tissues.
Keywords :
Three-dimensional printing , Knee Joint , Anatomical models , Anterior cruciate ligament reconstruction , Total knee arthroplasty , CAD