Title of article
Neck fracture of the Exeter stem in 3 patients
Author/Authors
Reito, Aleksi Coxa Hospital for Joint Replacement, Tampere, Finland , eskelinen, Antti Coxa Hospital for Joint Replacement, Tampere, Finland , PAJAmäki, orma Coxa Hospital for Joint Replacement, Tampere, Finland , PuolAkkA, timo Coxa Hospital for Joint Replacement, Tampere, Finland
Pages
4
From page
193
To page
196
Abstract
A 63-year-old woman (height 1.70 m, weight 92 kg, BMI 31)
with developmental dysplasia of both hips underwent bilateral
hybrid total hip replacement (THR) using the cemented Exeter
V40 stem (offset 44 mm, size 3) and uncemented Trident PSL
(Stryker) with alumina-on-alumina ceramic bearing (Biolox
Forte; Ceramtec, Plochingen, Baden-Württemberg, Ger-
many) at our hospital in 2004. A ceramic femoral head with
a diameter of 36 mm and offset of +5 mm was used. Short-
term functional outcome was excellent. 5 years and 3 months
later, the patient tripped on a flat floor and fell down, and was
unable to move her left leg due to severe pain. Plain radio-
graphs revealed a neck fracture in the Exeter stem. She under-
went revision surgery, during which both the remnant of the
Exeter stem and the uncemented acetabular component were
found to be well fixed.
Keywords
Neck fracture , Exeter stem , patients
Journal title
Acta Orthopaedica
Serial Year
2016
Record number
2618091
Link To Document