• Title of article

    Determination of the Shape and Dimensions of the Sensitive Volume for Solid State Detectors Using Monte Carlo Computer Technique

  • Author/Authors

    kadhim, alaa b. university of baghdad - college of science - department of astronomy and space, Iraq , mahdi, shams d. university of baghdad - college of science - department of astronomy and space, Iraq

  • From page
    1982
  • To page
    1991
  • Abstract
    In this research the active volume for a number of solid-state detectors of the type of high-purity germanium (HPGe) crystal was evaluated with different radii and depths using scanning method for diagonal (front) and lateral (side). It has been used for this purpose Monte-Carlo efficiency program after its development by adding a subroutine-program for its (subroutine scanning).Also a program has been written to calculate the stopping power and range for incident charged particle on the detector, in order to determine the exact sufficient energy to stop it inside the detector material. The calculations of our results of efficiency were compared with the results of published efficiency and the comparison is very good in terms of improving the values for practical from the calculated results in other ways.
  • Keywords
    Efficiency , Active Volume (Side and Front) , High , Purity Germanium (HPGe) Detector , Scanning , Solid , State Detectors and Monte , Carlo Method
  • Journal title
    Iraqi Journal Of Science
  • Journal title
    Iraqi Journal Of Science
  • Record number