New mode for the on-line determination of oxonium ion in different strong acids using CFIA via the use of homemade linear array Ayah 5Sx4-ST- 5D solar CFI analyser
Author/Authors :
shakir, issam m. a. university of baghdad - college of science - department of chemistry, Iraq , al-razack, hussein f. a. university of baghdad - college of science - department of chemistry, Iraq
From page :
282
To page :
302
Abstract :
Strong acids were determined via the blue complex solution reaction of a loaded ferric (III) ion on strong cation exchange resin which in turn reacts with potassium hyxacyano ferrate (II). The attenuation effect of formed blue complex Fe4([Fe(CN)6])3 on (0 -180°) incident LED light was measurement via homemade linear array AYAH 5SX4-ST-5D solar CFI analyser. Optimum parameters were 0.005M.L^-1 [Fe(CN)6]^-4 , flow rate of 3mL.min^-1 , sample volume 204 μL , sample purge time of 70 seconds was chosen, and 1V for light intensity. A liner calibration graph of 0.005 -0.15 M.L-1 were obtains for HCl, HNO3, HCLO4 and H2SO4, with a linearity (r^2 %) 94 -98 % and L.O.D based on gradual dilution of lowest concentration in calibration graph was 22.314 μg for HCl, 38.563 μg for HNO3, 61.480 μg for HCLO4, and 60.020 μg for H2SO4 .Repeatability (RSD) 0.5% was obtained. the proposed method was compared with standard method. turbidity measured using HANNA instrument for available acid in commercial ( HCl, H2SO4).
Keywords :
Turbidity , Oxonium ion , ion exchange