• Title of article

    Detection of methicillin-resistant Staphylococcus aureus directly by loop-mediated isothermal amplification and direct cefoxitin disk diffusion tests

  • Author/Authors

    Metwally, L. Suez Canal University - Faculty of Medicine - Department of Clinical Pathology, Egypt , Gomaa, N. Suez Canal University - Faculty of Medicine - Department of Clinical Pathology, Egypt , Hassan, R. Suez Canal University - Faculty of Medicine - Department of Clinical Pathology, Egypt

  • From page
    273
  • To page
    279
  • Abstract
    We evaluated the utility of 2 methods for detection of methicillin-resistant Staphylococcus aureus (MRSA) directly from signal-positive blood culture bottles: loop-mediated isothermal amplification (LAMP) assay, and direct cefoxitin disk diffusion (DCDD) test using a 30 μg cefoxitin disk. In parallel, standard microbiological identification and oxacillin susceptibility testing with MecA PCR was performed. Of 60 blood cultures positive for Gram-positive cocci in clusters, LAMP (via detection of the FemA and MecA genes) showed 100% sensitivity and specificity for identification of MRSA/MSSA. When coagulase-negative staphylococci were tested, sensitivity for detection of methicillin resistance was 91.7% and specificity was 100%. DCDD along with direct tube coagulase assay detected only 80.6% of MRSA/MSSA. LAMP showed higher diagnostic accuracy although DCDD was more cost-effective and did not require additional reagents or supplies.
  • Journal title
    Eastern Mediterranean Health Journal
  • Journal title
    Eastern Mediterranean Health Journal
  • Record number

    2644508