• Title of article

    OBJECTIVE QUALITY CONTROL PROGRAM FOR MEDICAL ULTRASOUND SCANNING

  • Author/Authors

    sharawy, amr a. cairo university - biomedical engineering department, Giza, Egypt , mohammed, kamel. k. cairo university - biomedical engineering department, Giza, Egypt , aouf, mohamed higher technological institute - biomedical engineering department, Tenth of Ramadan City, Egypt

  • From page
    1357
  • To page
    1361
  • Abstract
    Ultrasound imaging of a focal lesion of low contrast against background tissue necessitates maximum spatial and contrast sensitivity resulting in non-specular reflection, predisposing these images to degradation due to speckle. In recent years, diagnostic ultrasound image quality has greatly improved due to advances in system hardware and software. Developments in transducer design have resulted in transducers with greater bandwidth and sensitivity. Quality features to be tested are spatial resolution, contrast resolution and lesion signal-to-noise ratio. These features are explained and examples are shown.
  • Keywords
    Contrast , Resolution , Speckle Noise , Phantom , Ultrasound Imaging
  • Journal title
    Journal of Al Azhar University Engineering Sector (JAUES)
  • Journal title
    Journal of Al Azhar University Engineering Sector (JAUES)
  • Record number

    2649830