Title of article :
A Novel Method for Pulsometry Based on Traditional Iranian Medicine
Author/Authors :
Yousefipoor, Farzane tarbiat modares university - Department of Electrical and Computer Engineering, تهران, ايران , Nafisi, Vahidreza Iranian Research Organization for Science and Technology - Department of Electrical and Information Technology, ايران
From page :
230
To page :
237
Abstract :
Arterial pulse measurement is one of the most important methods for evaluation of healthy conditions. In traditional Iranian medicine (TIM), physician may detect radial pulse by holding four fingers on the patient’s wrist. By using this method, under standard condition, the detected pulses are subjective and erroneous, in case of weak and/or abnormal pulses, the ambiguity of diagnosis may rise. In this paper, we present an equipment which is designed and implemented for automation of traditional pulse detection method. By this novel system, the developed noninvasive diagnostic method and database based on the TIM are way forward to apply traditional medicine and diagnose patients with present technology. The accuracy for period measuring is 76% and systolic peak is 72%.
Keywords :
Automation , Heart Rate , Pulse , Wrist , Fingers , Cardiovascular Physiological Phenomena , Traditional Iranian Medicine