Average Absorbed Breast Dose (2ABD) to Mean Glandular Dose (MGD) Conversion Function for Digital Breast Tomosynthesis: A New Approach
Author/Authors :
Tucciariello, Raffaele M Medical Physics Unit - St.Chiara University Hospital, Pisa, Italy , Lamastra, Rocco Medical Physics Unit - St.Chiara University Hospital, Pisa, Italy , Barca, Patrizio Department of Medical Physics - St. Orsola-Malpighi University Hospital, Bologna, Italy , Evelina Fantacci, Marine Department of Physics - University of Pisa, Roma, Italy , Traino, Antonio C National Institute for Nuclear Physics (INFN), Pisa Section, taly
Pages :
9
From page :
183
To page :
191
Abstract :
Background: In this work a new method for the Mean Glandular Dose evaluation in digital breast tomosynthesis (DBT) is presented. Methods: Starting from the experimental-based dosimetric index, 2ABD, which represents the average absorbed breast dose, the mean glandular dose MGD2ABD was calculated using a conversion function of glandularity f(G), obtained through the use of Monte Carlo simulations. Results: f(G) was computed for a 4.5 cm thick breast: from its value MGD2ABD for different compressed breast thicknesses and glandularities was obtained. The comparison between MGD2ABD estimates and the dosimetric index provided in the current dosimetry protocols, following the Dance's approach, MGD , showed a good agreement (<10%) for all the analyzed breast Dance thicknesses and glandularities. Conclusion: The strength of the proposed method can be considered an accurate mean glandular dose assessment starting from few and accessible parameters, reported in the header DICOM of each DBT exam.
Keywords :
Digital breast tomosynthesis , breast dosimetry , absorbed breast dose
Journal title :
Archive of Breast Cancer
Serial Year :
2021
Record number :
2700236
Link To Document :
بازگشت