• Title of article

    Thermoluminescent Characteristics of GR-200, TLD-700H and TLD-100 for Low Dose Measurement: Linearity, Repeatability, Dose Rate and Photon Energy Dependence

  • Author/Authors

    Alizadeh ، Mojgan Department of Medical Physics and Radiology - Faculty of Paramedical Sciences - Kashan University of Medical Sciences , Mohseni ، Mehran Department of Medical Physics and Radiology - Faculty of Paramedical Sciences - Kashan University of Medical Sciences , Farhood ، Bagher Department of Medical Physics and Radiology - Faculty of Paramedical Sciences - Kashan University of Medical Sciences , Aliasgharzadeh ، Akbar Department of Medical Physics and Radiology - Faculty of Paramedical Sciences - Kashan University of Medical Sciences , Moradi ، Habiballah Department of Medical Physics and Radiology - Faculty of Paramedical Sciences - Kashan University of Medical Sciences , Ramazani-Moghaddam-Arani ، Ahmad Faculty of Physics - University of Kashan

  • From page
    111
  • To page
    116
  • Abstract
    Background: The dose values obtained from procedures of diagnostic radiology are relatively low. To accurately and precisely measure the dose values in this dose range, it is necessary to know the characteristics of dosimeters. Objective: The aim of this study was to evaluate several thermoluminescent characteristics of GR-200, TLD-700H and TLD-100 for low dose measurement. Material and Methods: In this experimental study, linearity, repeatability, dose rate and photon energy dependence of different TLD materials were investigated in a 0.05-10 mGy range dose. It is noteworthy that the data obtained from TLD-100 were considered as reference and the data obtained from two other types of TLDs were compared with them. Results: For all three types of TLD materials, there are linear relations between absorbed dose values to TLDs and their responses. TLD-100 and TLD-700H have very low sensitivity than GR-200. For GR-200 and TLD-100, the coefficients of variation values (%) are 3.00% and 2.01%, respectively, that these values are within the tolerance limit ( 7.5%). However, this value for TLD-700H is 10.85% which it is more than the reported tolerance limit. Furthermore, remarkable effects of dose rate and photon energy dependence on the responses of GR-200 are not observed in a 0.5-4 mGy dose range; nevertheless, remarkable effects of dose rate and photon energy dependence on the responses of TLD-100 and TLD- 700H are found in this dose range. Conclusion: The evaluated thermoluminescent characteristics for GR-200 are better than two other types of TLDs (TLD-100 and TLD-700H) for low dose values.
  • Keywords
    Thermoluminescent Dosimetry , Radiology , Radiation Dosage , GR , 200 , TLD , 100 , TLD , 700H
  • Journal title
    Journal of Biomedical Physics and Engineering
  • Journal title
    Journal of Biomedical Physics and Engineering
  • Record number

    2706964